How to install a heat flux sensor

Tips and tricks to get the most out of your heat flux measurement

Measuring heat flux is a powerful tool to gain insights in processes. You may measure for example how much heat flows through a wall, or to a specimen that must be cooled. Assuming the right sensor is used, installing this sensor correctly, so that it performs a stable measurement and measures the right heat flux (radiative and convective), is a critical step to get the right data. This paper dives into the do’s and don’ts when installing a heat flux sensor.

Read the full article here: How to install a heat flux sensor and get the most out of your heat flux measurement (PDF)

Introduction

Heat flux sensors have a wide variety of applications, from thermal performance analysis of thermal insulation, to monitoring of fouling of pipelines and the health monitoring of pigs. Measuring the heat flux can lead to useful insights in processes and system performance. Assuming the right sensor is used, mounting this sensor correctly, so that it performs a stable measurement and measures the right heat flux (radiative and convective), is a critical step to get the right data.

This paper focuses on sensor installation. What are the do’s and don’ts when installing a heat flux sensor; how can you get the best data from your sensor.

General considerations for heat flux measurement

  • use the right sensor for the application. There are many different models each with its own temperature- and heat flux range. View our complete product range of heat flux sensors.
  • see also our video on YouTube: how to measure heat flux.
  • perform a representative measurement. This starts with choosing the right location, representative for the system to be monitored. Use multiple sensors. The representativeness may be reviewed using infrared cameras.

Considerations for installation

Regardless of the heat flux sensor type, it is important that it is mounted securely in order to avoid variations of contact resistance between the sensor and the object on which it is mounted.

  • air gaps between sensor and object may be significant thermal resistances and increase response time. This should be avoided.
  • sensors gradually getting loose essentially produce unreliable (apparently unstable) measurements. Use a stable glue or filler. Use high quality cabling and strain relief.

Also, optical properties must match.

  • pay attention to the optical properties of the sensor surface. These must match those of the object the sensor is mounted on.

Mounting

There are various ways to mount a heat flux sensor, depending on the application. Two important parameters are

  • temperature range
  • the duration of the measurement

These two parameters will help choosing the right mounting solution for the heat flux sensor. Table 1 and the examples at the end of this note will help you review your options.

Always ensure strain relief on the cable to avoid unnecessary stress on the sensor.

Why to avoid air gaps

The thermal conductivity of air is in the order of 0.02 W/(m·K). Therefore, even small air gaps are significant thermal resistances.

The thermal conductivity of a plastic or thermal paste is in the order of 0.2 W/(m·K), so for the same thickness, thermal resistance is a factor 10 lower.

Take for example a 0.05 x 10⁻³ m, air gap. This has a thermal resistance of 20 x 10⁻⁴ K/(W/m²). This may be compared to 11 x 10-⁴ K/(W/m²) for FHF05 series or 70 x 10-⁴ K/(W/m²) for HFP01, so a small air gap produces an increase of thermal resistance of respectively 200 % for FHF and about 30 % for HFP01. Using a filler of 0.05 x 10⁻³ m, with a thermal conductivity around 10 times higher than that of air, the thermal resistance is reduced to 2.5 x 10-⁴ K/(W/m²). The contribution the thermal resistance reduces to about 20 % for FHF05 and 3 % for HFP01.

From this example you can also see that it is not necessary to use high-thermal conductivity tapes. Using a thin normal tape is enough.

An air gap may not only lead to a higher thermal resistance for conductive heat, but also to an entirely different radiation balance. An air gap is a “resistance” (a radiation screen) for radiative transfer. If it is filled-up, it is no resistance any longer. Watch out in case radiative (far infra-red) heat flux is significant. In that case the presence of an air gap may be the dominant source of errors, because a sensor with an air gap acts as a radiation shield, reducing local radiative transfer by a theoretical maximum of 50 %.

What to do about air gaps

Tapes, sheet (gasket) material, glues and cements al may be used to fill up air gaps.

These gaps may occur:

  • because of the nature of the surface. It may not be smooth. Smoothen before installation
  • because of a curvature in the surface. For all practical purposes a surface with a radius of smaller than 5 m is considered “flat”. At smaller radii, use of flexible sensors may be considered. For industrial sensors like IHF01 and IHF02, we may also provide coupling pieces (flat on one side, curved on the other).

Table 1 summarises the different mounting options.

Table 1 Options for mounting heat flux sensors. Materials may act to fix the sensor position, but also to fill up airgaps.

NRproductdurationrated temperature rangefunctionalitycomments
[#][description][description][˚C][description][description]
1powerstriptemporary, easily removable15 to 40fixation and gap fillingTESA Powerstrip. very easily removable.
2glycerineminutesto 120gap filling onlyFiller only for quick experiments; glycerine can be obtained at the local pharmacy. It is safe to use and easily dissolves in water.
3toothpastedays40gap filling only

Filler only, use with other fixation such as single sided tape

Water-based

Most commercially available toothpastes are suitable

4double sided tape2 weeks, removable40fixation and gap fillingTESA 4939 floor laying (carpet) tape combines a high initial bonding power with a residue free removability up to 14 days from the most common surfaces. (needs to be tested individually before usage)
5thermal pasteweeksto 177gap filling only

Filler only, use with other fixation such as single sided tape

Silicone oil-based

DOW CORNING heat sink compound 340

OMEGATHERM conductive paste

6silicone gluepermanent-45 to 200fixation and gap filling

Most commercially available silicone glues are suitable

DOWSIL 3145 silicone sealant

7single sided tapetemporary or permanent-260 to 150fixation only

Fixation only, use with other fillers such as thermal paste

TESA 51408 orange masking tape

Most commercially available Kapton tapes are suitable

NRproductdurationrated temperature rangefunctionalitycomments
[#][description][description][˚C][description][description]
8magnetstemporary or permanentto 500fixation only

on magnetic surfaces only

for sensors with optional “frame with magnet” only

in case using welded treads or bolting is not

9tack welded threadstemporary or permanent-260 to 1000fixation only

For sensors with flanges

Fixation only, use with other fillers such as silicone, graphite sheet material or cements

Usually combined with springs

10boltstemporary or permanent-260 to 1000fixation only

For sensors with flanges

Fixation only, use with other fillers such as silicone, graphite sheet material or cements

Usually combined with springs

11silicone gaskettemporary or permanentto 200Gap filling only

Filler only, use with other fixation such as bolts or threads

ERIKS silicone sheet 0.5 mm

Users can cut sheets to size

12graphite gaskettemporary or permanentto 500Gap filling only

Filler only, use with other fixation such as bolts or threads

ERIKS Ergaflex or similar sheet material

Users can cut sheets to size.

13high temperature cementtemporary or permanentto 1400fixation and gap filling

OMEGA high temperature cement

 

Other options for mounting

 

     
14Cements and epoxiesvariousvariousvariousOMEGA cements and epoxies

 

Why optical properties are important

When heat flux sensors are mounted at a surface, heat will often be transferred by a combination of radiation and convection. For the convective part, the thermal resistance of the sensor should be as low as possible. For the radiative part, the optical surface properties of the sensor should be representative of the surrounding area.

Some points to keep in mind:

  • radiation is not only transmitted in the spectral range that humans can see (visible radiation) but also as non-visible far infra-red
  • blank metal is reflective in the visible as well as in the far infra-red
  • paints and plastic coatings wood and stone absorb in different ranges, depending on their colour in the visible range. These materials typically all behave as “black” in the far infra-red. See figure 2.

The representativeness may be reviewed using a combination of normal (visible range) and infrared (far infra-red range) cameras.

Read the full article here: How to install a heat flux sensor and get the most out of your heat flux measurement (PDF)